FV1200 Lasers and Fluorescence Filters

Lasers

This microscope is equipped with a pulsed 405 diode, 559 DPSS, 440 & 635 diodes, and an argon laser with 458, 488, 515 lines.

Detectors

2x Spectral detectors - The spectral detectors on this microscope has a range of 400nm to 800nm.  The width can be set between 1nm and 100nm.

1x PMT

2x HSD - High sensitivity GaAsp Detectors.  

Filter cubes in the microscope

 

These filters are only used when you view your specimen down the microscope eyepieces. The filter turret will automatically rotate to Position 1 (LSM) when you start scanning with the confocal.

Position NumberFilter SetExcitation FilterDichroic MirrorEmission FilterExample Fluorophores
1LSMN/AN/AN/AEmpty - for confocal imaging
2U-FUW340 - 390410420IFDAPI, Hoechst
3U-FBNA470 - 495505510IFFITC, Alexa 488
4U-FGWA530 - 550570575IFTRITC, Texas Red
5U-FCFP425 - 445CFP455CFP460 - 510CFPCFP
6U-FGFP460 - 480GFP490CFP495 - 540GFPGFP
7U-FFN/AN/AN/AEmpty
8IX3-FDICTN/AN/AN/ADIC Analyser

Excitation Dichroics

Position NumberExcitation DichroicsExample Fluorophores
1BS 20/80 
2DM 405/488DAPI, FITC
3NoneEmpty
4DM 405/488/559/635DAPI, FITC, TRITC, Cy5
5DM 458/515CFP, YFP
6NoneEmpty

 

Emission Filters in the light paths

Filters in CHS1, CHS2 & CHS3 light path

DetectorDichroic MirrorFunctionWavelengths passedExample FluorophoresNotes
CHS1MirrorReflects all light to CHS1  See details of spectral detectors above
 GlassPasses all light through to other detectors   
 SDM560Directs light below 560nm to CHS1, and the rest to the other detectorsBelow 560nmAcridine Orange 
 SDM510Directs light below 510nm to CHS1, and the rest to the other detectors
Below 510nmCFP 
 SDM490Directs light below 490nm to CHS1, and the rest to the other detectors
Below 490nmDapi 
CHS2MirrorReflects all light to CHS2   
 GlassPasses all light through to other detectors   
 SDM640Allows light below 640nm to CHS2, and the rest to the other detectorsBelow 640nmTRITC, Texas Red 
 SDM560Allows light below 560nm to CHS2, and the rest to the other detectorsBelow 560nmGFP, FITC, AF488 
CH3MirrorReflects all light to CH3  This is a PMT with band pass filters before the detector
 GlassPasses all light through to HSD detectors   
 SDM640Allows light below 640nm to CH3. Passes light above 640nm to HSDsBelow 640nm There are three band pass filters available for CH3.  A 655-755, 575-675 and a 575-620.

Filters in HSD1&2 scanhead

Dichroic mirrors splitting light between HSD1 & 2

(These dichoics send the shorter wavelengths to HSD1 and longer wavelengths to HSD2

  • DM505
  • DM570
  • DM485
  • SDM510
  • SDM570
  • SDM640

Band pass filters to HSD1

  • BA460-600
  • BA495-540HQ
  • BA420-460
  • BA465-485
  • BA480-495
  • BA505-540
  • BA490-540
  • BA575-620

Band pass filters to HSD2

  • BA520-560
  • BA575-630
  • BA495-540HQ
  • BA605-680
  • BA535-565
  • BA575-675
  • BA655-755